Is an ISO 4406 Result reliable?

 

In the following test we seek to define the best ISO 4406
sampling methodology; using 5 analysis methods
.
 

  • Online accountants
  • Portable laboratory equipment
  • Stationary laboratory equipment
  • Laboratory microscope
  • Portable microscope
OFFLINE MEASUREMENT WITH CLEANING PROCESS
*|END:WEB_VIDEO|*
ONLINE MEASUREMENT
*|END:WEB_VIDEO|*

Measuring Highly High Codes


Using recirculation equipment equipped with 2 online ISO 4406 laser
particle  counters, with  a 0.83 gpm  pump and a beta 4>4193  filter,
the reservoir was filled in recirculation mode without filtration.

An ISO code 24/23/18 could be determined, which was ratified in the
other devices without any problem

 

Conclusion: The veracity of the results was 100% reliable.

Measuring Clean Codes

The fluid was filtered below <9<8<7, and the sample was taken through the counter. The surprise was that despite having repeated the procedure 3 times trying to clean the other equipment as much as possible, we were never able to obtain measurements below 16/14/12; 19/17/13 being the mean of the results of the first measurement procedure.
Conclusion: it was determined that the process of taking the sample and the same laboratory equipment contributed contamination with high results. 